  "case_id": "legacy-06-client_t761_eakin_sw_weekly_checkin_new",
  "transcript": "transcript.docx",
  "gold": "gold-minutes.md",
  "legacy_oracle": "legacy-expected.json",
  "origin": "existing",
  "meeting_type": "Technical file review",
  "difficulty": "existing",
  "required_actions": [
    {
      "id": "legacy-a1",
      "owner": "Rebecca",
      "action": "Add clarity to the risk management plan around probability-table number justification and what counts as an event.",
      "due": null
    },
    {
      "id": "legacy-a2",
      "owner": "David",
      "action": "Add a front page to the software-list Excel explaining what each tab represents and the purpose of the file.",
      "due": null
    },
    {
      "id": "legacy-a3",
      "owner": "Jacqui",
      "action": "Check whether the real-time clock battery alarm issue is captured in the risk analysis.",
      "due": null
    },
    {
      "id": "legacy-a4",
      "owner": "Andrew",
      "action": "Confirm the nebulizer flow-rate specification so David and Colm can assess ISO 27427 applicability.",
      "due": null
    },
    {
      "id": "legacy-a5",
      "owner": "David and Colm",
      "action": "Review ISO 27427 again once Andrew confirms the nebulizer flow-rate specification.",
      "due": null
    },
    {
      "id": "legacy-a6",
      "owner": "David",
      "action": "Send additional debug command letters to Andrew.",
      "due": null
    },
    {
      "id": "legacy-a7",
      "owner": "Andrew",
      "action": "Test the debug commands and report what happens on the debug screen.",
      "due": null
    },
    {
      "id": "legacy-a8",
      "owner": "Andrew",
      "action": "Resolve the remaining language character and font-driver issues, including access to the font-creator tool.",
      "due": null
    },
    {
      "id": "legacy-a9",
      "owner": "Andrew",
      "action": "Complete the IEC 60601-1 versus MDD documentation review and define the remaining electrical compliance testing.",
      "due": null
    },
    {
      "id": "legacy-a10",
      "owner": "David",
      "action": "Review the alarm code changes and later repeat the review for the language-selection software changes.",
      "due": null
    },
    {
      "id": "legacy-a11",
      "owner": "David",
      "action": "Prioritise tracing the 17 software changes from version 1.01 to 1.02 and identify any retrospective test-data needs.",
      "due": null
    },
    {
      "id": "legacy-a12",
      "owner": "Rebecca",
      "action": "Update the risk files, plan and matrix with the agreed cybersecurity and USB-port-lock approach.",
      "due": null
    }
  ],
  "tests": [
    "legacy_regression"
  ]
}